Quality control of instruments for measuring the characteristics of bactericidal UV radiation

https://doi.org/10.32362/2500-316X-2024-12-1-123-132

Abstract

Objectives. Quality control of instruments for measuring bactericidal irradiance of ultraviolet (UV) radiation is based on studying the main metrological characteristics. These characteristics include: angular and spectral sensitivity; linearity range; and absolute calibration in irradiance units. Deviations of the angular sensitivity of measuring instruments from the ideal cosine characteristic can significantly impact error estimation. They can also lead to the distortion of measurement results and a significant difference in instrument readings. The aim of this work is to enhance accuracy in resolving metrological problems of determining irradiance of bactericidal radiation.

Methods. An effective method of resolving this problem is to introduce correction coefficients for the angular sensitivity of radiometers, spectroradiometers and dosimeters. The values are calculated based on the results of measurements on the goniometer when testing measuring instruments. An important role is played by computer models and digital twins of measuring instruments based on the results of studies of the metrological characteristics of radiometers by means of software. This includes modeling the measuring task.

Results. The study of angular dependence of bactericidal UV radiometer sensitivity complemented by an analysis of measurement results obtained by other authors allows determining the value of the angular sensitivity correction coefficients by the deviation of the angular sensitivity of the irradiance measuring instruments of bactericidal radiation from the standard cosine dependence.

Conclusions. Deviations of the angular dependence of bactericidal radiation UV radiometer sensitivity from the cosine characteristic lead to a significant underestimation of the irradiance measurements results from extended emitters. An effective solution is the use of digital angular sensitivity correction coefficients to measure the irradiance of bactericidal radiation determined during tests. When assessing the quality of radiometers, spectroradiometers and dosimeters for bactericidal radiation, incomplete control of the main metrological characteristics of the measuring instruments creates risks of serious errors in the measurement results of bactericidal irradiance.

About the Authors

O. A. Minaeva
MIREA – Russian Technological University
Russian Federation

Olga A. Minaeva - Dr. Sci. (Eng.), Head of the Department of Metrology and Standardization, Institute for Advanced Technologies and Industrial Programming, Scopus Author ID 6603019847.

78, Vernadskogo pr., Moscow, 119454


Competing Interests:

The authors declare no conflicts of interest



A. V. Ragutkin
MIREA – Russian Technological University
Russian Federation

Alexander V. Ragutkin - Cand. Sci. (Eng.), Rector’s Advisor, Scopus Author ID 56871217700, ResearcherID AAE-4437-2022.

78, Vernadskogo pr., Moscow, 119454


Competing Interests:

The authors declare no conflicts of interest



S. I. Anevsky
MIREA – Russian Technological University
Russian Federation

Sergei I. Anevsky - Dr. Sci. (Eng.), Professor, Department of Metrology and Standardization, Institute for Advanced Technologies and Industrial Programming.

78, Vernadskogo pr., Moscow, 119454


Competing Interests:

The authors declare no conflicts of interest



R. V. Minaev
Elektrosteklo
Russian Federation

Roman V. Minaev - Cand. Sci. (Eng.), Head of the Research Department, Scopus Author ID 22235214600.

5, 26 Bakinskikh Komissarov ul., Moscow, 119571


Competing Interests:

The authors declare no conflicts of interest



Th. T. Minh
MIREA – Russian Technological University
Russian Federation

Thanh T. Minh - Postgraduate Student, Department of Metrology and Standardization, Institute for Advanced Technologies and Industrial Programming.

78, Vernadskogo pr., Moscow, 119454


Competing Interests:

The authors declare no conflicts of interest



I. A. Romanova
MIREA – Russian Technological University
Russian Federation

Inna A. Romanova - Senior Lecturer, Department of Metrology and Standardization, Institute for Advanced Technologies and Industrial Programming.

78, Vernadskogo pr., Moscow, 119454


Competing Interests:

The authors declare no conflicts of interest



References

1. Kostyuchenko S.V., Tkachev A.A., Frolikova T.N. UV technologies for disinfection of water, air and surfaces: principles and possibilities. Epidemiologiya i Vaktsinoprofilaktika = Epidemiology and Vaccination Prevention. 2020;19(5):112–119 (in Russ.). https://doi.org/10.31631/2073-3046-2020-19-5-112-119

2. Kostyuchenko S.V., Volkov S.V., Tkachev A.A., Strelkov A.K., Smirnov A.D., Baranov V.L. Modern trends and peculiarties of UV disinfection usage in water treatment. Vodosnabzhenie i sanitarnaya tekhnika = Water Supply and Sanitary Equipment. 2022;7:4–11 (in Russ.). https://doi.org/10.35776/VST.2022.07.01

3. Blatchley E.R. III, Brenner D.J., Claus H., Cowan T.E., Linden K.G., Liu Y., Mao T., et al. Far UV-C radiation: An emerging tool for pandemic control. Critical Reviews in Environmental Science and Technology. 2023;53(6):733–753. https://doi.org/10.1080/10643389.2022.2084315

4. Endo T., Gemma A., Mitsuyoshi R., et al. Discussion on effect of material on UV reflection and its disinfection with focus on Japanese Stucco for interior wall. Sci. Rep. 2021;11(1):21840. https://doi.org/10.1038/s41598-021-01315-1

5. Neu D.T., Mead K.R., McClelland T.L., Lindsley W.G., Martin S.B., Heil G., See M., Feng H.A. Surface Dosimetry of Ultraviolet Germicidal Irradiation Using a Colorimetric Technique. Ann. Work Expo. Health. 2021;65(5):605–611. https://doi.org/10.1093/annweh/wxaa147

6. Zong Y., Brown S.W., Johnson B.C., Lykke K.R., Ohno Y. Simple spectral stray light correction method for array spectroradiometers. Appl. Opt. 2006;45(6):1111–1119. https://doi.org/10.1364/AO.45.001111

7. Larason T.C., Cromer C.L. Sources of Error in UV Radiation Measurements. J. Res. Nat. Inst. Stand. Technol. 2001;106(4): 649–656. https://doi.org/10.6028/jres.106.030

8. Vijeta Kapri R.K., Saha S., Jaiswal V.K., Sharma P. Theoretical Simulation for Evaluating Error in Irradiance Measurement Using Optical Detectors Having Different Cosine Responses. MAPAN – Journal of Metrology Society of India. 2021;36(3): 473–480. https://doi.org/10.1007/s12647-021-00486-6

9. Marquez J.M.A., Bohórquez M.A.M., Garcia J.M., Nieto F.J.A. A New Automatic System for Angular Measurement and Calibration in Radiometric Instruments. Sensors. 2010;10(4):3703–3717. https://doi.org/10.3390/s100403703

10. Anevskii S.I., Zolotarevskii Yu.M., Ivanov V.S., Krutikov V.N., Minaeva O.A., Minaev R.V. Spectroradiometry of ultraviolet radiation. Meas. Tech. 2016;28(11):1216–1222. http://doi.org/10.1007/s11018-016-0873-9 [Original Russian Text: Anevskii S.I., Zolotarevskii Yu.M., Ivanov V.S., Krutikov V.N., Minaeva O.A., Minaev R.V. Spectroradiometry of ultraviolet radiation. Izmeritel’naya Tekhnika. 2015;11:26–30 (in Russ.).]

11. Mekaoui S., Zibordi G. Cosine error for a class of hyperspectral irradiance sensors. Metrologia. 2013;50(187):187–199. http://doi.org/10.1088/0026-1394/50/3/187

12. Nevas S., Wübbeler G., Sperling A., Elster C., Teuber A. Simultaneous correction of bandpass and stray-light effects in array spectroradiometer data. Metrologia. 2012;49(2):S43–S47. http://doi.org/10.1088/0026-1394/49/2/S43


Supplementary files

1. General view of a goniometer designed for measuring angular sensitivity of bactericidal UV radiation radiometers
Subject
Type Исследовательские инструменты
View (217KB)    
Indexing metadata ▾
  • Deviations of the angular dependence of bactericidal radiation UV radiometer sensitivity from the cosine characteristic lead to a significant underestimation of the irradiance measurements results from extended emitters.
  • An effective solution is the use of digital angular sensitivity correction coefficients to measure the irradiance of bactericidal radiation determined during tests.